2009 - 2013 cases, 2014 more cases will be uploaded soon.

Patient #197 (3 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#8, 9, 12

Immediate placement.

 

2013 09/18

8, 9- 5.0 x 10mm implant with 5.1mm healing abutment

12- 3.5 x 12mm implant with 4.2mm healing abutment

2013 12/06

3 months

Patient #192 (5 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes - 2

#5, 7, 10, 11, 13

13- Healed site.

5, 7, 10, 11- Immediate placement.

 

2013 08/09

5- 5.0 x 12mm implant with 4.6mm Closed Tray Abutment

7, 10- 4.3 x 10mm implant with 4.2mm Closed Tray Abutment

11- 4.3 x 12mm implant with 4.6mm Closed Tray Abutment

13- 4.3 x 10mm implant with 4.6mm Closed Tray Abutment

2014 05/02

9 months

2013-11-12

#5 and 7 implants failed after immediate temporary bridge fractured in #10 area after about 2 months in function. Pt. fractured bridge for two weeks prior to coming to the office. At that point implants 5 and 7 supported bridge 5-9. Implants did not have a chance...

Implants were removed, sites debrided (no buccal bone damage or infection was present).

See case #202 for replacement of these implants as well as additional #3 implant placement.

Patient #191 (6 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#19, 29, 24, 25, 29, 30

20, 29- Healed sites

19, 24, 25, 30- Immediate placement

 

2013 07/29

19, 30- 5.0 x 8mm implant with 5.0 x 6.5mm prepable abutment

20- 5.0 x 10mm implant with 4.6 x 6.5mm prepable abutment

24, 25- 2.5 x 14mm tom-KAT implant with 4.2 x 6.5mm contoured abutment

29- 4.3 x 10mm implant with 4.6 x 6.5mm prepable abutment

 

A temporary bridge was fabricated the same visit

2014 05/02

10 months

Patient #190 (5 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#3, 4, 5, 8, 9

3, 4, 5- Healed site

8, 9-  Immediate placement

 

2013 07/18

3- 3.5 x 8mm implant with 4.2 x 6.5mm prepable abutment

4- 2.5 x 10mm tom-KAT implant with 4.2 x 6.5mm prepable abutment

5- 2.5 x 12mm tom-KAT implant with 4.2 x 6.5mm prepable abutment

8, 9- 4.3 x 10mm implant with 4.6 x 10° angled abutment

 

A temporary bridge was fabricated the same visit.

2014 03/14

8 months

Patient #178 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#12

Immediate placement.

 

2013 05/21

5.0 x 10mm implant with 4.2mm temporary abutment crown.

A temporary crown was fabricated the same visit.

2013 07/01

2 months

Patient #168 (7 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes 1

#3, 4, 5, 8, 9, 13, 14

3, 4, 13- Healed site.

5, 8, 9, 14- Immediate placement.

 

2013 03/20

3, 14- 5.0 x 6mm implant

4- 3.5 x 8mm implant with 4.2mm closed tray abutment.

5- 4.3 x 10mm implant with 4.6mm closed tray abutment.

8, 9- 5.0 x 12mm implant with 5.4mm closed tray abutment.

13- 3.5 x 12mm implant with 4.6mm closed tray abutment.

A temporary bridge was fabricated and cemented over implants with temporary cement around the margins the same visit.

2013 06/03

3 months

#13 was removed and replaced on 06/19/2013 due to 2mm bone loss. #13 was not placed deep enough (was placed in a healed site), excessive bone loss happened because of bone remodeling around #14 implant (placed in a fresh extraction socket). Implant was unscrewed and new implant of the same dimensions was placed 2-3mm deeper. No infection was present. (see case #184)

Patient #165 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement

 

2013 03/08

3.0 x 12mm implant with 4.2mm temporary abutment.

A temporary crown was fabricated the same visit.

2013 06/06

3 months

Patient #145 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#13

Immediate placement

 

2012 10/22

5.0 x 10mm implant with 4.6 x 8.5mm prepable abutment

A temporary crown was fabricated the same visit.

2013 07/22

9 months

Patient #136 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement

 

2012 08/08

3.5 x 12mm implant with 4.2mm temporary abutment crown

A temporary crown was fabricated the same visit.

2013 01/14

5 months

Patient #134 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes

#5

Immediate placement

 

2012 07/26

4.3 x 12mm implant with 4.6 x 8.5mm prepable abutment.

A temporary crown was fabricated the same visit.

2012 08/17

3 weeks

#5 implant was removed on 08/17/2012 due to bone loss around the implant. A new implant was placed on 01/28/2013 (see case #160).

Extracted tooth had horizontal root fracture and large amount of mobility, might be the reason for failure.

Patient #133 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#29

Immediate placement

 

2012 07/17

5.0 x 8mm implant with 4.6 x 8.5mm prepable abutment.

A temporary crown was fabricated the same visit.

 

2014 08/12

2 years

Patient #131 (4 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes 1

#3, 4, 5, 6

Immediate placement

 

2012 07/02

3- 8.0 x 6mm implant with 6.4 x 8.5mm prepable abutment.

4- 5.0 x 10mm implant with 4.6 x 8.5mm prepable abutment.

5, 6- 4.3 x 12mm implant with 4.6 x 8.5mm prepable abutment.

A temporary bridge was fabricated the same visit.

 

2014 07/17

2 years

 

Note #3 implant was removed and replaced. Image shown reflects new implant (see case #172)

#3 implant was removed on 11/13/2012 due to poor soft tissue response. The implant was fully integrated. A new implant was placed on 4/9/2013 (see case #172)

Patient #130 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#12

Immediate placement

 

2012 06/27

5.0 x 10mm implant with 4.6 x 8.5mm contoured abutment.

A temporary crown was fabricated the same visit.

2012 09/06

3 months

Patient #126 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#25

Immediate placement

 

2012 05/18

3.0 x 14mm One-Piece implant

2013 02/08

1 year

Patient #122 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement

 

2012 05/04

3.5 x 14mm implant with 4.2mm prepable abutment

A temporary crown was fabricated the same visit.

2012 11/13

6 months

Patient #115 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement

 

2012 03/16

3.5 x 14mm implant with 4.2mm temporary abutment crown.

A temporary crown was fabricated the same visit.

2014 03/18

2 years

Patient #113 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#8, 9

Immediate placement

 

2012 03/08

8- 3.5 x 12mm implant with 4.6mm closed tray abutment

9- 3.5 x 14mm implant with 4.6mm closed tray abutment

Temporary crowns were fabricated the same visit.

2012 12/11

9 months

Patient #107 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5, 6

5- Healed site

6- Immediate placement

 

2012 01/10

5- 2.5 x 12mm implant with 4.7mm healing abutment.

6- 3.5 x 14mm implant with 4.2mm healing abutment.

2014 09/18

3 years

Patient #102 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#8

Immediate placement.

 

2011 12/06

4.3 x 14mm implant with 5.1mm temporary abutment.

A temporary crown was fabricated the same visit.

2014 05/20

2.5 years

Patient #101 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5

Immediate placement.

Implants #4 and 5 were removed at this visit due to lack of attached gingivae and thin buccal bone in #4 that resulted in gray color shining through. Both implants were removed and connective tissue grafting was done in #4, 5 area.

 

2011 11/04

3.0 x 12mm implant with 4.2mm temporary abutment.

A temporary crown was fabricated the same visit.

2014 03/03

2.5 years

Patient #99 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#9

Immediate placement.

 

2011 11/02

4.3 x 12mm implant with 5.1mm temporary abutment.

A temporary crown was fabricated the same visit.

2013 03/21

1.5 years

Patient #96 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes

#7

Immediate placement.

 

2011 10/26

3.5 x 14mm implant with old revision 4.2mm temporary abutment, crown. A temporary crown was fabricated the same visit.

2012 02/09

Implant was removed d/t infection. Buccal and lingual bone was missing.

2012 08/08

Grafting was done to correct buccal/lingual bone bone defect.

2012 11/19

PA post removal of titanium mesh.

Patient #90 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement.

 

2011 09/21

3.0 x 12mm implant with 4.2mm temporary abutment. A temporary crown was fabricated the same visit.

2012 11/28

1 year

Patient #79 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5

Bone did not heal properly around this implant (see Pat. #46 for initial implant placement).

 

2011 06/03

5.0 x 10mm implant

2012 08/27

1 year

 

Patient #73 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#9, 12

9- Immediate placement

12- Healed site

 

2011 04/04

9- 4.3 x 12mm implant

12- 3.5 x 10mm implant

Both with old style temporary abutments and temporary crowns fabricated same visit.

2012 04/02

1 year

Patient #71 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#10

Immediate placement.

 

2011 03/17

3.5 x 12mm implant with old style temporary abutment.  A temporary crown was fabricated the same visit.

2011 09/22

6 month

Patient #67 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7, 8

Immediate placement.

 

2011 02/15

7 & 8- 3.5 x 14mm implant with old style temporary abutment. Temporary bridge was fabricated the same visit.

2014 03/25

3 years

Patient #65 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement.

 

2011 02/15

4.3 x 14mm implant with old style temporary abutment. A temporary crown was fabricated the same visit.

2013 04/24

2 years

Patient #64 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#10

Immediate placement.

 

2011 02/15

4.3 x 10mm implant was placed with old revision temporary abutment. A temporary crown was fabricated the same visit.

2013 07/25

2.5 years

Patient #59 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#10

Immediate placement.

 

2011 01/17

3.5 x 14mm implant was placed with old style temporary abutment. A temporary crown was fabricated the same visit.

 

2013 02/18

2 years

Patient #46 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes

#5

Immediate placement.

 

2010 11/01

4.3 x 8mm implant was placed with old style temporary abutment.  A temporary crown fabricated the same visit.

#5 implant was removed on June 3, 2011. Bone did not heal properly after extraction. No infection was present, the implant was integrated. Implant was replaced the same visit with a 5.0 x 10mm implant.

Patient #40 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5

Immediate placement.

 

2010 10/07

5.0 x 10mm implant was placed with old style temporary abutment.  A temporary crown fabricated the same visit.

2013 08/14

3 years

Patient #27 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#20

Immediate placement.

 

2010 07/12

5.0x8mm implant was placed and a temporary crown was made over the old revision temporary abutment at same visit.

2011 11/03

~1.5 years

Patient #13 (1 implant)

Pre-op x-ray

Post-op x-ray

Most recent x-ray

Failure - No

#4

Immediate placement.

 

2010 03/11

5.0 x 8mm implant, old design with thread/fins combination. 4.7mm old design healing abutment was placed same visit. Immediate temporization with old style temp. abutment.

2014 04/30

4 years

 