Patient #6 (1 implant)

Pre-op x-ray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Healed site.

 

2009 11/23

2.5 x 12mm implant, old design. 4.2mm old design healing abutment was placed same visit.

2014 11/20

5 years

 

Patient #16 (2 implants)

Pre-op x-ray

Post-op x-ray

Most recent x-ray

Failure - No

#30, 31

Healed site.

 

2010 03/29

5.0 x 8mm implant, old design with thread/fins combination. Implant was placed too close to #11. Implant integrated well but had to be removed. See case # 68 for replacement implant.

 2012 09/11

2.5 years

 

Patient #26 (4 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#25, 26, 27, 28

Healed site, implants and #25 were removed ~5 months prior (implants had failed due to overloading by partial, area was debrided and bone grafting was done).

 

2010 07/08

25- 2.5 x 14mm implant, old design with thread/fins combination. 4.2mm old design healing abutment placed.

26- 3.5x12mm implant, old design with thread/fins combination. 4.2mm old design healing abutment placed.

27-  3.5x12mm implant, old design with thread/fins combination. 4.7mm old design healing abutment placed.

28- 4.3 x 8mm implant, old design with thread/fins combination. 4.7mm old design healing abutment placed.

2011 07/21

1 year

Patient #33 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#4, 5

5 was a healed site and 4 was immediate placement.

 

2010 09/08

4- 3.0 x 12mm implant was placed.

5- 4.3 x 8mm implant was placed.

2014 03/03

4 years

#4, 5

Both #4 and 5 implants integrated well but were removed 11/28/2011 because of aesthetic concerns (lack of keratinized gingiva and thin buccal bone plate)

#5 was replaced same visit (see case #101)

#4 was replaced 02/08/2012 (see case #112)

Patient #34 ( 2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#20, 28

Healed site.

 

2010 09/08

20- 5.0 x 8mm implant was placed.

28- 2.5 x 12mm implant was placed with 4.2mm healing abutment.

2014 06/16

~4 years

Patient #35 ( 2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#30, 31

Healed sites.

 

2010 09/09

30 & 31- 2.5 x 10mm implants were placed.

2013 06/18

~3 years

Patient #43 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#20, 21

Healed sites.

 

2010 10/25

20- 4.3 x 6mm implant was placed with  4.7mm healing abutment.

21- 3.0 x 12mm implant was placed with  4.7mm healing abutment.

2014 12/11

4 years

Patient #47 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#12

Healed site.

 

2010 11/04

3.0 x 12mm implant was placed with 4.7mm healing abutment.

2011 04/28

5 months

Patient #49 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Healed site.

 

2010 11/19

3.0 x 14mm implant was placed with old style temporary abutment.  A temporary crown was fabricated the same visit.

2013 04/29

2.5 years

Patient #51 (5 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5, 6, 7, 11, 12

5 & 6- Immediate placement.

7, 11, 12- Healed sites.

 

2010 12/13

5- 4.3 x 10mm implant,  5.1mm healing abutment

6- 4.3 x 12mm implant, 4.7mm healing abutment

7- 3.0 x 14mm implant, 4.2mm healing abutment

11- 3.5 x 14mm implant, 4.7mm healing abutment

12- 4.3 x 8mm implant, 4.7mm healing abutment

2014 02/10

3 years

Patient #54 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes

#23

Healed site.

 

2011 01/07

2.5 x 14mm implant was placed with old style 4.2mm temporary abutment.  A temporary crown was fabricated the same visit.

2011 05/23

5 months

Bone loss due to implant surface contamination with temporary crown material (observed after implant removal). Lesson learned - make larger removal screw excess hole to allow cement to escape. Place enough temporary resin to establish contacts with adjacent teeth and more.

Patient #75 (4 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#6, 7, 10, 11

Healed sites.

 

2011 04/21

6, 7, 11- 3.5 x 14mm implant with 4.7mm healing abutment.

10- 3.0 x 14mm implant with 4.7mm healing abutment.

2011 08/16

4 months

Pt. is out of state, was restored by another dentist. Working on getting the current x-rays.

Patient #77 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#26, 27

Healed sites.

 

2011 06/06

26- 3.0 x 14mm implant

27- 4.3 x 10mm implant

2014 09/11

3 years

Patient #82 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#23, 26

Healed sites.

 

2011 07/13

23- 3.0 x 14mm implant with 4.2mm healing abutment.

26- 2.5 x 14mm implant with 4.2mm healing abutment.

2013 01/03

1.5 years

Patient #100 (4 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#21, 23, 26, 28

Immediate placement.

 

2011 11/04

21, 28- 4.3 x 8mm implant with 4.2mm healing abutment.

23, 26- 3.0 x 14mm implant with 4.2mm healing abutment.

2014 04/28

2.5 years

Patient #101 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5

Immediate placement.

Implants #4 and 5 were removed at this visit due to lack of attached gingivae.

 

2011 11/04

3.0 x 12mm implant with 4.2mm temporary abutment.

A temporary crown was fabricated the same visit.

2014 03/03

2.5 years

Patient #107 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#5, 6

5- Healed site

6- Immediate placement

 

2012 01/10

5- 2.5 x 12mm implant with 4.7mm healing abutment.

6- 3.5 x 14mm implant with 4.2mm healing abutment.

2014 09/18

3 years

Patient #112 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#4

Healed site.

 

2012 02/08

3.0 x 12mm implant with 4.2mm healing abutment.

2014 03/03

2 years

Patient #120 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#26

Healed site

 

2012 04/20

3.0 x 14mm One-Piece Implant

Pt. moved to another state, trying to get an x-ray

Pt. has not returned after the implant and temp. crown placement. Likely due to financial issues.

Patient #125 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - Yes 1

#7, 10

7- 3.5 x 10

10- 2.5 x 10 for use as tenting screw

 

2012 05/17

7- 3.5 x 10mm implant

10- 2.5 x 10mm implant

2014 07/15

2 years

#7

Both #7 and 10 sites had non-KAT implants that failed resulting in a large bone defects without buccal or palatal bone remaining. Allograft bone blocks were used to restore these defects, implants were placed several months later. Failure of implant #10 is the result of poor integration of the block graft in that site.

Patient #126 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#25

Immediate placement

 

2012 05/18

3.0 x 14mm One-Piece implant

2013 02/08

1 year

Patient #128 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7, 10

Healed sites

 

2012 05/24

7, 10- 3.0 x 14mm implant with 4.2mm healing abutment

2014 06/16

2 years

Patient #132 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#4, 5

Healed sites

 

2012 07/17

4- 3.5 x 8mm implant with 4.2 x 8.5mm contoured abutment.

5- 2.5 x 12mm implant with 4.2 x 8.5mm contoured abutment.

A temporary bridge was fabricated the same visit.

2015 02/17

2.5 years

Patient #137 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#25

Healed site

 

2012 08/17

2.5 x 14mm One-piece implant

2012 10/11

2 months

Patient #141 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#22, 26

Immediate placement

 

2012 09/11

22- 3.5 x 12mm implant

26- 3.0 x 12mm implant

2014 11/17

2.5 years

Patient #147 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#25, 26

Healed sites

 

2012 10/30

25, 26- 2.5 x 14mm implant with 4.2 x 8.5mm prepable abutment

Temporary bridge was fabricated the same visit.

2013 12/19

1 year

Patient #165 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Immediate placement

 

2013 03/08

3.0 x 12mm implant with 4.2mm temporary abutment.

A temporary crown was fabricated the same visit.

2013 06/06

3 months

Patient #173 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7

Healed site.

 

2013 04/11

2.5 x 14mm implant with 4.2mm temporary abutment crown.

A temporary crown was fabricated the same visit.

2013 09/30

5 months

Patient #175 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#19

Healed site.

 

2013 04/16

3.0 x 12mm implant with 5.1mm healing abutment.

2014 01/27

9 months

Patient #177 (3 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#2, 3, 4

Healed sites.

 

2013 05/13

2- 5.0 x 6mm implant with 5.5mm healing abutment.

3- 4.3 x 10mm implant with 5.5mm healing abutment.

4- 3.0 x 10mm implant with 4.2mm healing abutment.

2013 11/11

6 months

Patient #182 (6 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#4, 5, 19, 20, 21, 22

Healed sites.

 

2013 06/10

4- 4.3 x 8mm implant with 4.7mm healing abutment

5- 4.3 x 10mm implant with 4.7mm healing abutment

19- 3.5 x 8mm implant with 4.7mm healing abutment

20- 3.1 x 14mm tom-KAT implant with 4.2mm healing abutment

21- 3.1 x 12mm tom-KAT implant with 4.2mm healing abutment

22- 3.1 x 14mm tom-KAT implant with 4.2mm healing abutment

2014 10/20

1.5 years

Patient #190 (5 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#3, 4, 5, 8, 9

3, 4, 5- Healed site

8, 9-  Immediate placement

 

2013 07/18

3- 3.5 x 8mm implant with 4.2 x 6.5mm prepable abutment

4- 2.5 x 10mm tom-KAT implant with 4.2 x 6.5mm prepable abutment

5- 2.5 x 12mm tom-KAT implant with 4.2 x 6.5mm prepable abutment

8, 9- 4.3 x 10mm implant with 4.6 x 10° angled abutment

 

A temporary bridge was fabricated the same visit.

2014 03/14

8 months

Patient #191 (6 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#19, 29, 24, 25, 29, 30

20, 29- Healed sites

19, 24, 25, 30- Immediate placement

 

2013 07/29

19, 30- 5.0 x 8mm implant with 5.0 x 6.5mm prepable abutment

20- 5.0 x 10mm implant with 4.6 x 6.5mm prepable abutment

24, 25- 2.5 x 14mm tom-KAT implant with 4.2 x 6.5mm contoured abutment

29- 4.3 x 10mm implant with 4.6 x 6.5mm prepable abutment

 

A temporary bridge was fabricated the same visit

2014 05/02

10 months

Patient #199 (2 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#22, 30

Healed sites.

 

2013 10/03

22- 3.0 x 14mm implant with 4.2mm healing abutment

30- 5.0 x 10mm implant

2014 11/08

1 year

Patient #205 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#30

Healed site.

 

2013 12/03

3.1 x 12mm implant

2014 05/05

5 months

Patient #220 (1 implant)

Pre-op xray

Post-op x-ray

Final restoration

Failure - No

#5 healed site after allograft cortical block graft.

 

 

2014 05/15

3.1 x 12mm

2015 05/07 cemented crown

Patient #225 (3 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#4, 6, 7

healed sites with infected #5 removal.

Immediate temporization.

 

2014 08/12

#4 - 5.0 x 10mm with Temp. Abutment Bridge 4.7

#6 - 3.5 x 14mm with Temp. Abutment Bridge 4.2

#7 - tom-KAT 3.1 x 14 with Temp. Abutment Bridge 4.2

 

Jet resin temporary bridge cemented with glass ionomer cement.

 

2014 02/24

4 months

Zirconia full contour bridge with Closed Tray Abutments. Closed tray impression technique.

Patient #232 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#24

healed site

Immediate temporization

 

2014 07/31

One-piece 2.5 x 14mm

Immediate temporization with flowable composite

 

 

2014 12/16

4 months

Conventional prep/impression

Patient #235 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#20

healed site, recent extraction, free soft tissue graft

 

2014 11/06

3.1 x 10mm

healing abutm. 4.2

thin ridge

 

2015 03/10

4 months

Closed tray abutm. 4.2

Closed tray impression

Full contour zirconia

 

Patient #238 (3 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

Healed sites. Pt. lost teeth 40+ years ago, extremely thin anterior ridge. pt. is on a budged, looking for a better denture fit.

 

2014-08-19

#2 - 5 x 6mm after internal sinus elevation

#10 - 2.5 x 12mm - the only anterior area it was possible to place the implant with bone expansion

#14 - 5 x 6mm after internal sinus elevation

 

 

2014 12/11

3 months

Patient #242 (3 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#11 immediate placement

#12 healed site

#13 placed same time with lateral sinus augmentation

 

2014 10/29

#11 - 2.5 x 12

#12 and 13 - 2.5 x 10

Thin ridge

Lateral sinus lift with allograft in #13 same visit

Flipper for a temporary

 

2015 01/13

3 months

Patient #245 (6 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

Immediate placement and loading of 6 implants in the mandible.

 

2014 11/14

#20 4.3 x 6

#21 4.3 x 6

#23 3.1 x 14

#27 3.5 x 12

#28 4.3 x 8

#29 4.3 x 8

2015 01/15

2 months

Patient #247 (1 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#10

Immediate placement

 

2014 11/18

3.1 x 14mm implant with 4.2mm healing abutment

 

Patient #250 (2 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7 and 10 healed sites. Thin bone, large undercuts. Immediate temporization.

 

2014 12/05

Immediate temporization

#7 and 10 - 2.5 x 14 implants, 4.2 temporary abutments, crown.

 

 

Patient #253 (4 implant)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

#7 immediate placement;

#3 and 13 healed sites, internal sinus lifts;

#10 healed site.

 

2014 12/12

#3 - 5 x 8

#7 - 3.5 x 12

#10 - 3.1 x 12

#13 - 4.3 x 6

No healing abutments, immediate denture delivered same visit.

 

 

Patient #256 (7 implants)

Pre-op xray

Post-op x-ray

Most recent x-ray

Failure - No

Immediate upper full arch placement and loading.

 

2015 01/09           Pt. would not consider the temporary denture.

#4-13 teeth were removed, implants were placed in same visit.

#4, 5 - 4.3 x 8 implant and Temporary Abutments 4.2; #6 - 4.3 x 10 implant and Temporary Abutments 4.2; #8 - 3.5 x 12 implant and Temporary Abutments 4.7; #10 - 3.1 x 14 implant and Temporary Abutments 4.2; #11 - 4.3 x 12 implant and Temporary Abutments 4.7; #12 - 5.0 x 8 implant and Temporary Abutments 4.7.

Buccal bone plate had green stick fracture during #4, 5 implants placement. Both #4 and #5 implants were inserted with 20 Ncm of torque only. Temporary bridge was fabricated to keep #4 and 5 out of bite.

#12 site had dense bone, implant was placed at about 50 Ncm. Temporary bridge was cemented with Ketac cement.

2014 09/22

7 months